This paper reports on investigations of air coupled ultrasonic imaging techniques which are part of the German research project MaTech. Project partners are Airbus, Bremen, Ingenieurbüro Dr. Hillger, Braunschweig, and IntelligeNDT, Erlangen, with DLR, Braunschweig, as a subcontractor.
One task is the development of a modular ultrasonic imaging system which opens the possibility of basic research of air-coupled ultrasonic techniques for composites.
The development and construction of special transducers for air-coupling and the development of a system for echo technique are further research tasks of the MaTech project. This paper presents the features of the demonstrator III and first results from investigations of transducers and their application to composites.
The air-coupled ultrasonic inspection is a very attractive method because it avoids the disadvantages of the coupling liquid (water, squirter-, immersion technique) or coupling paste [1]. However, the large impedance mismatch between solids and gas (air) produces an amplitude loss of more than 150 dB using through-transmission-technique with separate standard transducers as a transmitter and a receiver on opposite sides of a CFRP- component [2].
Therefore, a special equipment consisting of transducers with impedance-matching to air, a high-power pulser and an ultra-low noise preamplifier have to be used [3].
In order to obtain a high signal-to-noise ratio, resonant transducers are applied. Their low damping provides a high acoustic power output but a small bandwidth (about 10% of the centre frequency) .
Another method for increasing the signal to noise ratio is the application of narrow band filters in the receiver because the electronic noise is proportional to the square root of the bandwidth .
Narrow band transducers can produce resonances in the transmission coefficient. Fig. 1 shows the calculated l/2-resonances in dependence of the thicknesses of CFRP-plates (sound velocity of 3100 m/s) in through-transmission technique using a test frequency of 350 kHz. It is shown that a thickness of 4.4 mm (l/2) produces a 10 dB higher receiver amplitude than a 2 mm thickness. Because of the attenuation of 0,6 dB/mm the second resonance (2 l/2) delivers only a maximum of 4.5 dB. These effects have to be observed by an inspection with narrow band pulses. It is possible that areas with delaminations produce higher through-transmission amplitudes than those without defects.
Fig 1: Resonances in the through-transmission coefficient in dependence of the
thickness of CFRP-plates calculated for a test frequency of 350 kHz and an
attenuation coefficient of 0,6 dB/mm.
|
On the other hand, broadband air-coupled transducer are available; their smaller sensitivity can be compensated by pulse compression techniques [4].
There are several manufactures of air-coupled transducers. Usually the transducers are only used with the special equipment of the transducer manufacturer, so that it is difficult to compare their features and their technical data.
This paper reports on the aspect of „air-coupled ultrasonic technique“ of the German research project MaTech. It centres on the development of a fast imaging system (Demonstrator III), basic research, development of special transducers and on investigations of the detection of pores and delaminations in composites.
The demonstrator III provides basic research of air coupled ultrasonic technique and fast ultrasonic imaging. Fig. 1 shows the block diagram. The system is of modular design and enables different excitation pulses such as burst and free programmable signals. The special features are:
Fig 2: Block diagram of Demonstrator III.
|
Transducers from different manufacturers have been investigated within the frequency range of < 120 kHz to 1 MHz.
The transducers have been characterised in through-transmission technique in air. The transmitting transducer was excited by a rectangular pulse of Us = 200 V. The output from the receiving transducer was directly measured by an oscilloscope. The sensitivity was calculated by
where Ur is the voltage of the receiving transducer.
For the measurement of the frequency range of the transducers a spike-pulser based on avalanche-transistors was used. This kind of pulser produces a very broadband excitation. The received, amplified and gated signal was given to a spectrum-analyser. Tab. 1 gives an overview of three different transducers.
| Transducer | Nominal frequency kHz | Active diameter mm | Frequency
range kHz | Focus- distance mm | Beam- diameter mm | Focal Zone mm | Sensi-tivity |
| 1 | 120 | 19,3 | 115-135 | 30 | 13 | - | -32 dB |
| 2 | 250 | 50 | 232-247 | 50 | 1,4 | 10,3 | -43 dB |
| 3 | 1 MHz | 25 | 470-740 | 50 | 0,7 | 10,3 | -62 dB |
| Table 1: Transducers for air-coupled ultrasonic technique. | |||||||
The highest sensitivity of -32 dB was measured with 120 kHz-transducers (transducer 1). This transducer is useful for the inspection of sandwich-components with foam-core (see chapter 4). Transducer 2 with an active diameter of 50 mm produces a very small sound beam calculated as 1.4 mm. Note that the wavelength in air is also 1.4 mm using a frequency of 240 kHz!
Transducer 3 provides a large frequency range of 470 kHz to 740 kHz, measured with a distance of 25 mm (Fig. 3, left hand side). If the distance between the transducer is varying, the frequency spectrum is changed, at a distance of 66 mm it is divided in two parts (Fig. 3, right hand side).
Fig 3: Frequency spectra of transducer 3 measured in different distances.
|
The results of the measured sound beam diameter are shown in Fig. 4. Two transducers of the same type were used for these measurements. One transducer was fixed, the other one was scanned in different distances around the centre. In different distances, C-scans were recorded. The echo dynamics of the C-scans with the smallest beam diameter are shown in Fig. 4.
Fig 4: Profiles of the sound beams in
focal distances of different transducers.
|
Air-coupled transducers were used for ultrasonic imaging of defects in composites. Because the demonstrator III is still in development, the HFUS 2400 AirTech was used for the investigations.
Sandwich components with cores of foam or Nomex can only be penetrated with frequencies in a range of 50 to 800 kHz. Because of these low frequencies the ultrasonic wavelength is relatively large and reduces the resolution. Inspections are usually accomplished in through-transmission technique with separate transmitting and receiving transducers on opposite sides of the component (coupling with water, Squirter technique).
Fig. 5 shows a complex component with foam core and skins of CFRP. The component is stringer stiffened and was impacted with energies of 10 to 30 Joule at different positions (directly on the stringers and between two stringers). Therefore the sizes of the defects are quite different. The C-scan recorded in through-transmission technique with air-coupling (two transducers type 1) clearly shows the positions of the stringers with an amplitude of 0 dB, the areas with foam between the skins (-9 dB) and defect areas with amplitudes of - 22 to -24 dB.
Fig 5: Sandwich component with foam core, C-scan, construction (cross-section),
and echo-dynamic curve.
|
Fig 6: C-Scan and echo-dynamic curve of a CFRP- fabric sandwich component
with Nomex-cores (thickness of 21 mm).
|
Fig. 6 presents a C-scan of a sandwich with Nomex-cores and skins of CFRP. Two transducers 2 are used. The small beam diameter enables a clear presentation of the cores and of the defects. The echo dynamic curve clearly presents the different amplitude ranges for different areas: CFRP without cores deliver 0 dB , - 2 to –25 dB can be measured for defect-free areas with cores and –31 dB for damaged areas with cores.
This paper treats of ultrasonic investigations with air coupling which are part of the German research project MaTech.
Subtasks are: development and construction of an ultrasonic imaging system ("Demonstrator III") including special transducers, basic research of transducers, investigations of application to composites, and further developments of the potential of the echo-technique.
The demonstrator is under construction and will have the following features:
Fist investigations showed that :
Further investigations with free-programmable transmitter signals, signal analysis and multi-element transducers will be carried out for the application of echo-technique.